170 Ind. Admin. Code 5-3-2.2

Current through November 6, 2024
Section 170 IAC 5-3-2.2 - Calibration of instruments

Authority: IC 8-1-1-3; IC 8-1-22.5-4

Affected: IC 8-1-2; IC 8-1-22.5

Sec. 2.2.

(a) An instrument or tool, the use of which is necessary to comply with this rule, shall be calibrated according to the manufacturer's specifications and at intervals recommended by the manufacturer.
(b) Records of instrument calibration shall include the instrument serial number or unique identifier, date of calibration, the name and signature of the individual or third party vendor performing the calibration, the process and calibration standard used for the calibration, as-found and as-left data, and the due date for the next calibration.

170 IAC 5-3-2.2

Indiana Utility Regulatory Commission; 170 IAC 5-3-2.2; filed 3/17/2022, 12:34 p.m.: 20220413-IR-170210213FRA