Or. Admin. Code § 333-016-0035

Current through Register Vol. 63, No. 10, October 1, 2024
Section 333-016-0035 - Tentative Method of Test for Flashpoint of Volatile Flammable Materials by Tagliabue Open-Cup Apparatus*,**
(1) Scope:
(a) This method describes a test procedure for the determination of open-cup flashpoints of volatile flammable materials having flashpoints below 175 degrees Fahrenheit;
(b) This method, when applied to paints and resin solutions which tend to skin over or which are very viscous, gives less reproducible results than when applied to solvents.

*NOTE: The Public Health Division has obtained permission from the American Society for Testing Materials, Philadelphia, Pennsylvania, to reprint this method in their regulations. The test has been slightly modified for practical reasons.

**NOTE: ASTM Designation: D 1310-59T, issued 1954, revised 1955, 1956, and 1959. This tentative method has been approved by the sponsoring committee and accepted by the American Society for Testing Materials in accordance with established procedures for use pending adoption as standard. Suggestions for revisions should be addressed to the Society at 1916 Race Street, Philadelphia, Pennsylvania.

(2) Outline of Method. The sample is placed in the cup of a Tag Open Tester, and heated at a slow but constant rate. A small test flame is passed at a uniform rate across the cup at specified intervals. The flashpoint is taken as the lowest temperature at which application of the test flame causes the vapor at the surface of the liquid to flash; that is, ignite but not continue to burn.
(3) Apparatus. The Tag Open-Cup Tester is illustrated in Figure 1. It consists of the following parts which must conform to the dimensions shown and have the additional characteristics as noted:
(a) Copper bath, preferably equipped with a constant level overflow so placed as to maintain the bath liquid level one-eighth inch below the rim of the glass cup;
(b) Thermometer holder. Support firmly with ringstand and clamp;
(c) Thermometer. For flashpoints above 40 degrees Fahrenheit, use the ASTM Tag Closed Tester Thermometer, range of plus 20 to plus 230 degrees Fahrenheit, in one degree Fahrenheit divisions and conforming to thermometer 9F of ASTM Standard E1. For flashpoints from 20 degrees Fahrenheit to 40 degrees Fahrenheit, use ASTM Tag Closed Tester, Low Range Thermometer 57F. For flashpoints below 20 degrees Fahrenheit, use ASTM Thermometer 33F.;
(d) Glass test cup. Glass test cup (Figure 2) of molded clear glass, annealed, heat-resistant, and free from surface defects;
(e) Leveling device. Leveling device or guide, for proper adjustment of the liquid level in the cup (Figure 3). This shall be made of No. 18-gauge polished aluminum with a projection for adjusting the liquid level when the sample is added to exactly one-eighth inch below the level of the edge or rim of the cup;
(f) "Micro" or small gas burner of suitable dimensions for heating the bath. A screw clamp may be used to help regulate the gas. A small electric heater may be used;
(g) Ignition taper, which is a small, straight, blow-pipe type gas burner. The test flame torch prescribed in the method of test for flash and fire points by Cleveland Open Cup (ASTM designation: D 92) is satisfactory;
(h) Alternative methods for maintaining the ignition taper in a fixed horizontal plane above the liquid may be used, as follows:
(A) Guide wire, 3/32 inch in diameter and 3-1/2 inches in length, with a right-angle bend one-half inch from each end. This wire is placed snugly in holes drilled in the rim of the bath, so that the guide wire is 5/8 inch from the center of the cup and resting on the rim of the cup;
(B) Swivel-type taper holder, such as is used in ASTM Method D 92. The height and position of the taper are fixed by adjusting the holder on a suitable ringstand support adjacent to the flash cup.
(i) Draft shield, consisting of two rectangular sheets of noncombustible material, 24 inches x 28 inches, are fastened together along the 28 inch side, preferably by hinges. A triangular sheet, 24 inches x 24 inches x 34 inches is fastened by hinges to one of the lateral sheets (to form a top when shield is open). The interior of the draft shield shall be painted a flat black.
(4) Procedure:
(a) Place the tester on a solid table free of vibration, in a location free of perceptible draft, and in a dim light;
(b) Run water, brine, or water-glycol solution into the bath to a predetermined level, which will fill the bath to one-eighth inch below the top when the cup is in place. An overflow is permissible for water level control;
(c) Firmly support the thermometer vertically halfway between the center and the edge of the cup on a diameter at right angles to the guide wire, or on a diameter passing through the center of the cup and the pivot of the taper. Place so that the bottom of the bulb is one-fourth inch from the inner bottom surface of the cup. If the old Tagliabue thermometer is used, immerse to well cover the mercury bulb, but not the wide body of the thermometer;
(d) Fill the glass cup with the sample liquid to a depth just one-eighth inch below the edge, as determined by the leveling device;
(e) Place the guide wire or swivel device in position, and set the draft shield around the tester so that the sides form right angles with each other and the tester is well toward the back of the shield;
(f) If a guide wire is used, the taper, when passed, should rest lightly on the wire, with the end of the jet burner just clear of the edge of the guide wire. If the swivel-type holder is used, the horizontal and vertical positions of the jet are so adjusted that the jet passes on the circumference of a circle, having a radius of at least six inches across the center of the cup at right angles to the diameter passing through the thermometer, and in a plane one-eighth inch above the upper edge of the cup. The taper should be kept in the "off" position, at one end or the other of the swing, except when the flame is applied;
(g) Light the ignition flame and adjust it to form a flame of spherical form matching in size the 5/32-inch sphere on the apparatus;
(h) Adjust heater source under bath so that the temperature of the sample increases at a rate of two plus or minus 0.5 degrees Fahrenheit per minute. With viscous materials, this rate of heating cannot always be obtained.
(5) Initial Test. Determine an approximate flashpoint by passing the taper flame across the sample at intervals of two degrees Fahrenheit. Each pass must be in one direction only. The time required to pass the ignition flame across the surface of the sample should be one second. Remove bubbles from the surface of the sample liquid before starting a determination. Meticulous attention to all details relating to the taper, size of taper flame, and rate of passing the taper is necessary for good results. When determining the flashpoint of viscous liquids and those liquids that tend to form a film of polymer, etc., on the surface, the surface film should be disturbed mechanically each time before the taper flame is passed.
(6) Recorded Tests. Repeat the procedure by cooling a fresh portion of the sample, the glass cup, the bath solution, and the thermometer at least 20 degrees Fahrenheit below the approximate flashpoint. Resume heating and pass the taper flame across the sample at two intervals of five degrees Fahrenheit, and then at intervals of two degrees Fahrenheit until the flashpoint occurs.
(7) Reporting Data. The average of not less than three recorded tests, other than the initial test, shall be used in determining the flashpoint and flammability of the substance.
(8) Standardization:
(a) Make determinations in triplicate on the flashpoint of standard paraxylene and of standard isopropyl alcohol which meet the following specifications:
(A) Specifications of p-xylene, flashpoint check grade. P-xylene shall conform to the following requirements:
(i) Specific Gravity: 15.56 degrees Centigrade/15.56 degrees Centigrade, 0.860 minimum, 0.866 maximum;
(ii) Boiling Range: 2 degrees Centigrade maximum from start to dry point when tested in accordance with the method of test for distillation of industrial aromatic hydrocarbons (ASTM designation: D 850), or the method of test for distillation range of lacquer solvents and effluents (ASTM designation: D 1078). The range shall include the boiling point of pur-xylene, which is 138.35 degrees Centigrade (281.03 degrees Fahrenheit);
(iii) Purity: 95 percent minimum, calculated in accordance with the method of test for determination of purity from freezing points of high-purity compounds (ASTM designation: D 1016), from the experimentally determined freezing point, measured by the method of test for measurement of freezing points of high purity compounds for evaluation of purity (ASTM designation: D 1015).
(B) Specifications for isopropanol, flashpoint check grade. Isopropanol shall conform to the following requirements:
(i) Specific Gravity: 0.8175 to 0.8185 at 20 degrees Centigrade/20 degrees Centigrade, as determined by means of a calibrated pycnometer;
(ii) Distillation Range: Shall entirely distill within a 1.0 degree Centigrade range which shall include the temperature 80.4 degrees Centigrade as determined by ASTM method D 1078. Average these values for each compound. If the difference between the values for these two compounds is less than 15 degrees Fahrenheit (8.5 degrees Centigrade) or more than 27 degrees Fahrenheit (16 degrees Centigrade), repeat the determinations or obtain fresh standards.
(b) Calculate a correction factor as follows: X = 92 - A

Y = 71 - B

Correction:

X + Y/ Z

Where:

A = Observed flash of p-xylene

B = Observed flash of isopropyl alcohol

Apply this correction to all determinations. Half units in correction shall be discarded.

(9) Precision:
(a) For hydrocarbon solvents having flashpoints between 60 degrees Fahrenheit and 110 degrees Fahrenheit, repeatability is plus or minus two degrees Fahrenheit and the reproducibility is plus or minus five degrees Fahrenheit;
(b) If results from two tests differ by more than ten degrees Fahrenheit, they shall be considered uncertain and should be checked. The calibration procedure provided in this method will cancel out the effect of barometric pressure if calibration and tests are run at the same pressure. (Data supporting the precision are given in Appendix III of the 1956 Report of Committee D-1 on Paint, Varnish, Lacquers, and Related Products, Proceedings, American Soc. Testing Mats., Volume 56 (1956).)

Or. Admin. Code § 333-016-0035

HD 40, f. 8-2-73, ef. 8-15-73

The Figure(s) referenced in this rule is not printed in the OAR Compilation. Copies are available from theagency.

Publications: The publication(s) referred to or incorporated by reference in this rule are available from theagency.

Stat. Auth.: ORS 453.095

Stats. Implemented: ORS 453.005 & 453.055