Ex Parte Kantrowitz et alDownload PDFPatent Trial and Appeal BoardFeb 27, 201310770269 (P.T.A.B. Feb. 27, 2013) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. CONFIRMATION NO. 10/770,269 02/02/2004 Adrian Kantrowitz LTI-10302/38 9595 13173 7590 02/28/2013 Patent Procurement Services 450 North Old Woodward First Floor Birmingham, MI 48009 EXAMINER EVANISKO, GEORGE ROBERT ART UNIT PAPER NUMBER 3762 MAIL DATE DELIVERY MODE 02/28/2013 PAPER Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE ____________________ BEFORE THE PATENT TRIAL AND APPEAL BOARD ____________________ Ex parte ADRIAN KANTROWITZ and ROBERT MERRILL SMITH ____________________ Appeal 2011-002921 Application 10/770,269 Technology Center 3700 ____________________ Before: MICHAEL C. ASTORINO, REMY J. VANOPHEM, and NEIL T. POWELL, Administrative Patent Judges. POWELL, Administrative Patent Judge. DECISION ON APPEAL Appeal 2011-002921 Application 10/770,269 2 STATEMENT OF CASE Appellants appeal under 35 U.S.C. § 134 from a rejection of claims 1, 4, 5, 10-13, and 26-31. We have jurisdiction under 35 U.S.C. § 6(b). A hearing was held on Feb. 12, 2013. We AFFIRM. CLAIMED SUBJECT MATTER Independent claim 1, reproduced below, is illustrative of the appealed subject matter. 1. An implantable aortic blood pump, comprising: a wall having an outer surface, the wall adapted to maintain stability of the implantable aortic blood pump when implanted; a flexible membrane attached to the wall; an interior volume defined by the wall and the membrane; an inflation port housing integral with the wall, the inflation port housing having an aperture therein in fluid communication with the interior volume; a stiffener on the outer surface, or integral with the wall; and a suture flange extending from the outer surface, said suture flange adapted to secure to a human aorta, wherein the pump maintains geometry when implanted. REFERENCES The prior art relied upon by the Examiner in this appeal is: Kantrowitz 4,630,597 Dec. 23, 1986 Dobak 5,820,542 Oct. 13, 1998 Freed 6,511,412 B1 Jan. 28, 2003 Appeal 2011-002921 Application 10/770,269 3 REJECTIONS Appellants seek our review of the following rejections. 1. Claims 1, 4, 5, 10-13, 26, 27, and 29 stand rejected under 35 U.S.C. § 103(a) as obvious over Kantrowitz and Dobak. 2. Claim 28 stands rejected under 35 U.S.C. § 103(a) as obvious over Kantrowitz, Dobak, and Freed. 3. Claims 30 and 31 stand rejected under 35 U.S.C. § 103(a) as obvious over Kantrowitz and Dobak. ANALYSIS Rejection of Claims 1, 4, 5, 10-13, 16, 27, and 29 Appellants argue claims 1, 4, 5, 10-13, 26, 27, and 29 as a group. App. Br. 3 and 5. We select independent claim 1 as representative, and we treat claims 4, 5, 10-13, 26, 27, and 29 as standing or falling with representative claim 1. See 37 C.F.R. § 41.37(c)(1)(vii). The Examiner finds that Kantrowitz discloses a pump meeting most of the limitations of claim 1. Ans. 4. The Examiner finds, however, that “Kantrowitz does not disclose an integral stiffener or stiffener connected to the outer surface to maintain geometry of the pump when implanted.” Id. To address this, the Examiner finds that Dobak discloses using a stent with a pump to maintain pump geometry when implanted, and the Examiner concludes that it would have been obvious to modify Kantrowitz’s pump to include Dobak’s stent. Id. Appellants advance three arguments traversing the Examiner’s position. We consider each of Appellants’ arguments in turn below. Appeal 2011-002921 Application 10/770,269 4 Appellants’ Argument 1 Appellants argue that the Examiner’s proposed combination of the prior art does not meet claim 1 because “[a] stiffener cannot be equated with a stent.” App. Br. 6. In support of this, Appellants assert that the Markush group in original claim 3 set forth a “stent” and a “stiffener” as distinct elements of the species “element for stiffening,” and that the Specification similarly discloses a stent and a stiffener as distinct elements described in separate paragraphs. See Id. Additionally, Appellants argue that the term “stent” requires a collapsible configuration, whereas the term “stiffener” precludes a collapsible construction. Id. at 7. In support of this position, Appellants state that “in contrast to the collapsible nature taught by the specification as a characteristic of a stent, a stiffener is not taught to be collapsible.” Id. Appellants also argue that the collapsible nature of a stent conflicts with the claim language “wherein the pump maintains geometry when implanted.” Id. Appellants then assert that “[t]he three distinct elements available for stiffening an apparatus taught by the specification are separated by the unique and non-overlapping characteristics of each” and that any implication that the terms “stent” and “stiffener” overlap “represents an improper interpretation of pending claims in view of the guidance and definition provided in the specification.” Id. at 8 (emphasis added). The Examiner counters that the Specification does not define “stiffener.” Ans. 8. The Examiner also indicates that the term “stiffener” has a very broad ordinary meaning, and the Examiner notes that the Specification indicates that a “stent” falls within the scope of the similarly broad language “stiffening element.” Id. at 7-8. Regarding Appellants’ Appeal 2011-002921 Application 10/770,269 5 arguments about the collapsible nature of a stent, the Examiner states “the applicant has not claimed a non-collapsible stiffener.” Id. at 9. Regarding the related argument that the combination of Dobak’s stent with Kantrowitz’s pump would not meet the claim language “wherein the pump maintains geometry when implanted,” the Examiner indicates that the claim language does not clearly require that the pump maintain geometry at all times. Id. at 8. The Examiner further notes that the Specification suggests that the construction of the pump may allow it to deform (e.g., twisting and flexing) during installation and then maintain geometry after completion of installation. Id. In connection with this, the Examiner notes that “Dobak states that the purpose of the expansion mechanism/stent is to hold the housing in the expanded state and will retain its larger diameter until compressed by the application of fluid pressure (e.g. column 7, lines 19-37, column 8, etc. of Dobak).” Id. In response, Appellants reiterate the argument that original claim 3 and the Specification establish the term “stiffener” constitutes a species of the genus “element for stiffening.” Reply Br. 2. Regarding the temporal aspect of the claim language “wherein the pump maintains geometry when implanted,” Appellants assert that “[t]he language when implanted is an unambiguous statement of the amount of time, i.e. continually.” Id. at 1. Following this assertion, Appellants state that “the geometry is maintained during operation.” Id. at 2. Regarding the Examiner’s observation that the Specification discloses the pump may twist and flex during installation, Appellants point to the Freed Declaration made of record August 17, 2006. Id. Appellants argue that the Freed Declaration establishes that the pump disclosed in Kantrowitz has enough flexibility to allow folding, which Appeal 2011-002921 Application 10/770,269 6 Appellants assert constitutes more deformation than twisting and flexing. Id. To resolve the dispute regarding the proper interpretation of the term “stiffener” in claim 1, we begin with the disagreement regarding whether the Specification defines “stiffener.” Appellants do not point to and we do not find any explicit definition of “stiffener” in the Specification. Accordingly, we must consider whether the intrinsic evidence Appellants point to implicitly suggests a special definition of the term “stiffener.” As noted above, Appellants contend that the intrinsic evidence establishes a mutually exclusive relationship between the terms “stent” and “stiffener” because 1) the Specification and the original claims establish these terms both constitute species of the term “element for stiffening,” and 2) the Specification and claims establish that a “stent” has a collapsible configuration, whereas a “stiffener” has a configuration that cannot collapse. Appellants’ first point suffers from the problem that even if we accept that “stent” constitutes a species of “element for stiffening” and that “stiffener” also constitutes a species of “element for stiffening,” this does not preclude the possibility of a further species/genus relationship between “stent” and “stiffener” within the overarching genus “element for stiffening.” Compounding this problem, the intrinsic evidence relied on by Appellants state that the “element for stiffening” can include “a stent, a brace, a stiffener, or any combination thereof,” the last portion of which only serves to blur the relationship between the terms “stent” and “stiffener.” See Orig. Cl. 3 (emphasis added); Spec., p. 4, l. 20-p. 5, l. 2; p. 5, ll. 11-12. Further undermining Appellants’ position, original claim 3 and similar portions of the Specification indicate that a “stent” and a “stiffener” provide Appeal 2011-002921 Application 10/770,269 7 the same stiffening function. See also Spec., p. 4, l. 19-p. 5, l. 2; p. 5, ll. 10- 12. Thus, while the Specification’s general references to the terms suggest that “stent” and “stiffener” have different meanings with different scopes, they do not clearly establish that the terms have mutually exclusive scopes. This brings us to Appellants’ more specific contention that the intrinsic evidence establishes that the term “stent” requires a collapsible construction and the term “stiffener” precludes a collapsible construction. The Specification’s disclosure that the exemplary stent has a “collapsible” configuration does not clearly convey that a “stiffener” cannot have a collapsible configuration. Appellants correctly note that the Specification does not disclose that a “stiffener” has a collapsible configuration. But this silence also fails to clearly convey that a “stiffener” cannot have a collapsible configuration. Nor does any evidence pointed to by Appellants clearly suggest that collapsibility or the lack thereof constitutes a basis for distinguishing a “stent” from a “stiffener.” The evidence also fails to support Appellants’ argument that the claim language “wherein the pump maintains geometry when implanted” precludes the claimed “stiffener” from having a collapsible configuration. If we accept Appellants’ contention that this claim language necessarily has some limiting effect on the claimed “stiffener,” we must address the temporal aspect imparted by the words “when implanted.” These words could mean 1) during implantation, 2) after completion of implantation, or 3) both. Consistent with the Examiner’s findings, the intrinsic evidence repeatedly suggests a need to maintain pump geometry during operation, while also suggesting the acceptability and even desirability of allowing pump deformation during implantation. See, e.g., Spec., p. 18, ll. 8-12; p. Appeal 2011-002921 Application 10/770,269 8 13, ll. 13-15; and p. 3, l. 18-p. 4, l. 7; Reply Br. 2 (“As such, the geometry is maintained during operation.”); Ans. 9. Based on this, we conclude the claim requires that the pump maintains geometry after completion of implantation (i.e., during operation), but imposes no restriction on whether the pump maintains geometry during or prior to implantation. Thus, at most, the intrinsic evidence pointed to by Appellants may support a conclusion that the pump and the stiffener must maintain geometry after completion of implantation. Accordingly, while the intrinsic evidence may suggest one limitation on the definition of the term “stiffener,” it does not explicitly disclose or implicitly suggest a general definition that this limitation modifies. Where the Specification does not assign or suggest a particular definition for a claim term, it is appropriate to consult a general dictionary definition of the term for guidance in determining the ordinary and customary meaning thereof as viewed by a person of ordinary skill in the art. Comaper Corp. v. Antec, Inc., 596 F.3d 1343, 1348 (Fed. Cir. 2010) citing Phillips v. AWH Corp., 415 F.3d, 1303, 1322-23 (Fed. Cir. 2005). In the dictionary, the term “stiffener” is a noun associated with the verb “stiffen,” which means “[t]o make or become stiff or stiffer.” THE AMERICAN HERITAGE® DICTIONARY OF THE ENGLISH LANGUAGE (4th ed. 2000), accessed at http://www.thefreedictionary.com/stiffener (last visited December 26, 2012). This suggests that an ordinary meaning of the term “stiffener” is a structure that makes another structure stiff or stiffer. Taking this in combination with the above-discussed limitation suggested by the intrinsic evidence, we conclude that the term “stiffener” in claim 1 means a Appeal 2011-002921 Application 10/770,269 9 structure that makes the pump stiff or stiffer and that maintains geometry after completion of implantation. Given this, we consider next whether the Examiner correctly found that Dobak’s stent would constitute a “stiffener,” as recited in claim 1, when combined with Kantrowitz’s pump. As the Examiner notes, Dobak indicates that its stent has a configuration that allows it to hold a flexible housing “in an expanded substantially rigid state.” See, e.g., Dobak, Abstr; col. 6, ll. 57- 59; col. 7, ll. 4-37; Ans. 4. Thus, when added to Kantrowitz’s pump, Dobak’s stent would make the pump stiffer. As also noted by the Examiner, after it is expanded during implantation, Dobak’s stent holds its expanded shape until someone takes steps to collapse it. See, e.g., Dobak, col. 7, ll. 20-37; Ans. 4 and 8. Thus, when attached to Kantrowitz’s pump, Dobak’s stent would make the pump stiffer and would maintain pump geometry after completion of implantation of the pump and stent. Accordingly, the Examiner correctly found that Dobak’s stent falls within the broadest reasonable interpretation of the claimed “stiffener,” and that the combination of this stiffener with Kantrowitz’s pump would meet the claim language “wherein the pump maintains geometry when implanted.” Regarding Appellants’ arguments based on the Freed Declaration, even if accurate, these assertions establish only the flexibility of Kantrowitz’s pump without Dobak’s stent; they do not address whether the combination of Kantrowitz’s pump and Dobak’s stent would meet the claim limitations. And Appellants’ related argument regarding the “flexible nature of the stent of Dobak” (See Reply Br. 2) overlooks the Examiner’s correct finding regarding the disclosed rigidity of Dobak’s stent. Appeal 2011-002921 Application 10/770,269 10 Appellants’ Argument 2 Starting with the assumption that simply combining Dobak’s stent with Kantrowitz’s pump would not meet the limitations of claim 1, Appellants’ Argument 2 advances that modifying Dobak’s stent to meet the limitations of claim 1 would “destroy[] the function and principle of operation of the stent of Dobak.” App. Br. 8-9. As part of this argument, Appellants allude to the above-discussed argument that adding Dobak’s stent to Kantrowitz’s pump would not meet the claim limitations “wherein the pump maintains geometry when implanted.” Id. at 8 (Stating that “[i]t is noted that pending independent claims 1 and 5 recite that the pump maintains geometry when implanted, and for Dobak to do so would render Dobak inoperative to perform as intended.”). Appellants also assert that “[t]he claimed apparatus . . . requires rigidity to ‘maintain implant stability by resisting deformation.’” Id. at 9 (citing Spec., p. 14, ll. 19-20). In concert with this, Appellants suggest that Dobak’s collapsible configuration lacks the requisite rigidity. Id. The Examiner responds that Appellants’ err in suggesting that the combination of Dobak’s stent and Kantrowitz’s pump would not meet the claim language “wherein the pump maintains geometry when implanted.” See Ans. 9. And the Examiner asserts that the claims do not claim “rigidity to ‘maintain implant stability by resisting deformation,’” as suggested by Appellants. Id. The argument that modifying Dobak’s stent would destroy its function and change its principle of operation does not establish error in the rejection because Appellants have not established that one would need to modify Dobak’s stent to meet the limitations of claim 1. For the reasons discussed Appeal 2011-002921 Application 10/770,269 11 above in connection with Argument 1, Appellants have not established error in the Examiner’s position that Dobak’s stent corresponds to the claimed “stiffener” and that the combination of Dobak’s stent with Kantrowitz’s pump would result in a pump that meets the claim limitations regarding maintaining geometry when implanted. Regarding Appellants’ assertion that “[t]he claimed apparatus . . . requires rigidity to ‘maintain implant stability by resisting deformation,’” Appellants do not identify specific claim language that requires these features. Instead, Appellants point to the Specification in support of this proposition, and the portion of the Specification to which Appellants point does not even mention “rigidity.” See Spec., p. 14, ll. 19-20. We must decline to import these limitations from the Specification into the claim, particularly where the cited portion of the Specification itself does not clearly disclose the limitations. See In re Van Geuns, 988 F.2d 1181, 1184 (Fed. Cir. 1993) (Although the claims are interpreted in light of the specification, limitations from the specification are not read into the claims.) Appellants’ Argument 3 Appellants argue that “[t]he combination of Kantrowitz and Dobak . . . produce a pump that does not maintain geometry when implanted.” App. Br. 10. Appellants rest this position on arguments substantially the same as those presented in Argument 1, including arguments that Kantrowitz’s pump has a flexible construction, that Dobak’s stent has a collapsible construction, and that the combination would be collapsed during implantation. Id. at 10-11. The Examiner counters that the combination does meet the claim language “wherein the pump maintains geometry when Appeal 2011-002921 Application 10/770,269 12 implanted” for substantially the same reasons the Examiner explained in response to Appellants’ Argument 1. Ans. 9-10. As discussed above, Appellants’ arguments fail to recognize the temporal aspect of the claim language, as well Dobak’s disclosure that its stent holds its shape in the expanded state until someone takes active steps to collapse it. Again, because Dobak’s stent holds its shape in the expanded state, the combination of Dobak’s stent with Kantrowitz’s pump would allow the pump to maintain its geometry after completion of implantation. For the reasons discussed above, the claim requires nothing more. Accordingly, Appellants have not established error in the Examiner’s rejection of claim 1. Rejection of Claim 28 Appellants do not advance any separate arguments traversing the rejection of claim 28. See App. Br. 5-11. Accordingly, for the reasons discussed above, Appellants have not established error in the rejection of claim 28. Rejection of Claims 30 and 31 Appellants do not advance any separate arguments traversing the rejection of claims 30 and 31. See App. Br. 5-11. Accordingly, for the reasons discussed above, Appellants have not established error in the rejection of claims 30 and 31. Appeal 2011-002921 Application 10/770,269 13 DECISION We affirm the Examiner’s decision regarding claims 1, 4, 5, 10-13, and 26-31. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a)(1)(iv). AFFIRMED Klh Copy with citationCopy as parenthetical citation